indoor air quality meter to measure various parameters (CO2, CO, temperature, relative humidity) / calculates dew point, wet bulb temperature, absolute humidity, specific humidity, enthalpy, and atmospheric pressure)
The indoor air quality meter is a type of instrumentation that allows you to analyze and check air quality (INDOOR AIR QUALITY, IAQ). With the indoor air quality meter, you can measure various parameters CO2, CO, temperature, relative humidity, and it can also calculate the dew point, wet bulb temperature, absolute humidity, specific humidity, enthalpy, and atmospheric pressure using the P37AB147 probe. The only thing you cannot measure with this probe is carbon monoxide. This indoor air quality meter can also connect combined probes for measuring temperature and humidity, hot wire air velocity probes, fan air velocity probes, and other temperature probes. With the indoor air quality meter, you can calculate the percentage of outside air injection (% Outside Air) based on carbon dioxide, temperature, and renewal frequency. The indoor air quality meter has internal memory to store 67,600 values for each of the two inputs divided into 64 blocks. This indoor air quality tester has a graphical dot matrix display with a resolution of 160x160 dots according to the standard reference ASHRAE 62.1-2004, Italian Law Decree 81/2008 indicating that it applies to all enclosed spaces that can be occupied by people. Due to the presence of high humidity, kitchens, bathrooms, changing rooms, and swimming pools are also included. Any type of chemical, physical, and biological contaminant or outside air that is brought inside and not adequately purified must be taken into account according to air quality.
Applications for indoor air quality meter
- IAQ measurement and comfort conditions in schools, offices, and indoors
- Analysis and study of Sick Building Syndrome and its.
consecuencias
- Verification of HVAC system efficiency
- Examination of IAQ conditions in companies to optimize the microclimate and improve it
productividad
- Checks in Building Automation
Technical specifications
Carbon dioxide CO2
Sensor
Dual wavelength NDIR
Measurement range
0 ... 5000 ppm
Sensor operating range
-5 ... 50 °C
Accuracy
±50ppm+3% of the measurement
Resolution
1 ppm
Temperature dependence
0.1% f.s./°C
Response time (T90)
< 120 sec. (air speed = 2m/sec.)
Estabilidad a largo plazo
5% of the measurement/5 years
Carbon monoxide CO (only with the P37AB147 probe)
Sensor
Electrochemical cell
Measurement range
0 ... 5000 ppm
Sensor operating range
-5 ... 50 °C
Accuracy
±3ppm+3% of the measurement
Resolution
1 ppm
Response time (T90)
< 50 seg.
Vida aprox.
> 5 years under normal environmental conditions
Humedad relativa H.r.
Sensor type
Capacitivo
Sensor protection
Stainless steel network filter (on request, P6 filter of AISI316 sintered 20µm or P7 filter of sintered PTFE 10µm)
Measurement range
0 ... 100 % H.r.
Sensor operating range
-20 ... +60 °C
Accuracy
±2% (10÷90% H.r.) ±2.5% in the remaining field
Resolution
0,1 °C
Temperature dependence
+/-2% over the entire temperature range
Hysteresis and repeatability
1% H.r.
Response time (T90)
< 20 sec. (air speed = 2m/sec.) without filter
Estabilidad a largo plazo
1%/year
Temperatura T
Sensor type
Piezo-resistivo
Measurement range
750 ... 1100 hPa
Accuracy
±1.5 hPa @ 25°C
Resolution
1 hPa
Long-term stability
2 hPa/year
Temperature drift
±3 hPa with temperature -20 … +60 °C
Response time (T90)
< 20 sec. (air speed = 2m/sec.) without filter
Estabilidad a largo plazo
1%/year
Common technical specifications
Relative humidity
Sensor
Capacitivo
Capacidad típica @30% H.r.
300pF±40pF
Sensor operating temperature
-20 ... 80 °C
Measurement range
0÷100% H.r.
Incertidumbre
±1.5% RH (10…90% RH) ±2.5% RH in the remaining range
Resolution
0.1% H.r.
Deriva en temperatura @20°C
0.02%H.r./°C
Response time %H.r.
10 sec. (10÷80% H.r.; air speed=2m/s) at constant temperature
Temperature with Pt100 sensor
Resolution
0,1 °C
Deriva en temperatura @20 °C
0.003%/°C
Materiales
ABS, goma
Pantalla
Backlit, matrix with 160x160 dots, visible area 52x42 mm
Working conditions
Working temperature
Storage temperature
Relative humidity of operation
-5 ... 50 °C
-25 ... 65 °C
0 ... 85% H.r., no condensation
Type of protection
IP66
Incertidumbre
± 1 digito @ 20°C
Power supply
Power adapter (code SWD10)
Batteries
12Vdc/1A
4 x 1.2 V AA batteries
Autonomy
20 hours with alkaline batteries of 1800mAh (with P37AB147 probe inserted)
Interfaz
RS232C (EIA/TIA574) or USB 1.1 or 2.0 non-isolated
Memoria